Manual for pulse oximetry with PC |
|
The RS232 test kit has a SPO4025c OEM module with some extra circuit for working on a RS232 interface. In this case the power supply of the pulse oximeter and of the sensor is taken from the handshake lines of the serial interface. Please consider that on modern subnotebooks the RS232 interfaces are built such as to send just a minimum signal level. In this case it may be possible that not enough voltage is built up for the pulse oximeter. After connecting the test kits through a null modem cable the LED must blink with pulse rythm, if you put the sensor onto the finger. In problematic cases one can try it with a USB-to-RS232 adapter. We will also offer on the comercial market test kits with built in USB interfaces instead of RS232 interfaces. Then you don't need a separate USB-to-serial adapter and the null modem cable anymore. Before testing please download the
program CADT_Monitoring.exe. The application program opens the serial interface, so that the pulse oximeter gets voltage and starts. For that you choose in the menu of application program: Option -> Connection, the serial interface, where you connected the serial test kit. In case of a com interface it is normally COM1 oder COM2, but if it has a USB-to-serial adapter it is e.g. COM3 or COM4. A USB-to-serial adapter must be installed with the proper driver software, before starting our application program! The application program decodes the data stream from the oximeter module and shows the data in numerical way and as Plethysmogram. Red parts in the measurement curve show that the pulse oximeter adjusted the LED current or the preamplifier gain factor in one or more places. The plethysmogram curve must not show jumps, but continue as a smooth line. The black marks below the graph view the progress of the internal activities of the pulse oximeter module. E.g. the regular 0.1 second long marks indicate, when the pulse LED in the test kit was turned on. The tables show the set value and the measured value of the LED currents in the sensor. The measured LED current must match with the set value. A discrepancy of more than 10 % means a problem in the sensor. Using a common sensor with two LEDs means the third column of the table is irrelevant. In the third row we have the resulting
voltage of the photo
diode. For a good sensor
it is about 1,5 to 2
V, in case of a low light sensor it is
about
0,5 V. The fourth row shows under the inscription
"ADC Sig" the measured variation of the signal voltage. This
must not exceed more than 5 or 10 % of the signal. Otherwise
reliable pulse oximetry cannot be guaranteed. Reason for that can be
e.g. a defect sensor (e.g.
defect cable) or an interference from an active cellular phone
that is close to this place. The value with the inscription Gain indicates the effective transimpedance resistance of the photo current preamplifier. Below that there is a table row, which
shows RMS values of single samples (50 Hz) of the measured
photo signals. When you leave the sensor at rest for about 1 to 2
minutes, then these values indicate a measure for the noise of the
pulse oximeter. It should not exceed 0.02 %. Sensitivity for low
perfusion normally is a factor 2 to 3 lower, since a heart beat always
appears in many subsequent samples, i.e. we are reducing noise by
taking an average of several samples. Afterwards the table shows the pulse
oximetry ratings with a light blue inscription: perfusion, pulse, pulse rise time, pulse
jitter, saturation. In
addition
to the perfusion value we show the probability for the current
oximetric model in a range from 0 to 100. The measurement is considered "bad" , if
this probability was less the 30%. In
this case the previous results will be shown for a maximum
duration of 5 seconds
and afterwords they will disappear. Besides that, the
results don't
have any further dependence on the past. If the perfusion
level was determined to be less than 0.1 %, the other pulse oximetric
ratings will not be shown
on the display. The following table shows us 4 characteristic ratings:
If the pulse oximeter doesn't start, it can be a problem with the null modem cable. In this cable all the contacts must be connected, also the modem control lines. Due to the high data rate of 57,6 KBaud it is also important to activate the hardware FiFo buffers of the serial interfaces. Otherwise a successful data transfer is not possible under MS Win32. By the way, while running our application
program a pulse
oximetry record
file is written into
the file
fTxtLog.txt. You can save these files of format
"text with tabs" and use them with MS Excel. These files can
be very large. They will be overwritten once you start the application
program. Since May 2005 you can download the new Version 1.02. It contains these
enhancements: Copyright http://www.cadt.de |